Part No. 20717 Rev-01
2014-07
5-14
Low Priority Alarms
Alarm
Possible Cause
Recommended Action
20. Failed O
2
Sensor
Delivery of INOMAX continues during this
alarm.
a. O
2
calibration may have
drifted.
1. Perform low calibration.
2. Perform high O
2
calibration.
b. There is a leak around the
sensors.
Confirm the O-rings on the sensor and sensor
cover are correctly seated and the sensor cover
is fully closed.
c. O
2
sensor absent or failed.
If O
2
sensor is newly installed, allow 40 minutes
for conditioning of sensor.
1. Perform low calibration.
2. Perform high O
2
calibration.
Contact Technical Support.
21. Monitoring Failure
Delivery of INOMAX continues during this
alarm.
a. Monitor is failing to
communicate correctly or is
reporting a fault.
1. Confirm patient gas sample line is not
occluded.
2. Confirm water trap bottle and water separator
cartridge are in place.
3. Change patient gas sample line.
4. Change water separator cartridge.
Contact Technical Support.
22. Sample Line/Filter Block
All of these actions can be performed while
delivering INOMAX to the patient:
a. The sample line may be
blocked.
1. Confirm sample line inlet and outlet ports are
not obstructed.
2. Confirm patient gas sample line is not
occluded.
3. Change patient gas sample line.
4. Replace disk filter (if in use).
b. The water separator
cartridge may be blocked.
Replace water separator cartridge.
Contact Technical Support.
23. Two Cylinders Open
All of these actions can be performed while
delivering INOMAX to the patient:
a. Two cylinder valves are
open.
Close one INOMAX cylinder valve.
Contact Technical Support.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...