Part No. 20717 Rev-01
2014-07
6-4
Oxygen Sensor High Calibration
1. From the high calibration screen
(third menu level), press the
100% O
2
button to initiate the O
2
high calibration wizard.
3a. To continue the high calibration wizard
press the NEXT button.
3b. To start the high calibration without using
the wizard, press the START CAL button
3c. To exit the O
2
high calibration wizard, press
the CANCEL button.
2. Assemble connectors into a calibration setup
and set oxygen flow to five L/min.
(1) 100% O
2
Source
(2) O
2
Tubing
(3) 15M X 4.5 mm I.D Adapter
(4) Gas Sample Tee
(5) Patient Gas Sample Line with Nafion
1
2
5
3 4
3b
3a
3c
Caution:
• When performing a high calibration, make sure to select the correct calibration gas and
confirm the expiration date before using.
• Never connect the sample line directly to a high pressure gas source (greater than
150 cmH
2
O); this could damage the sampling system.
Note: Complete a low calibration (see Low Calibration section) prior to completing the high calibration.
The oxygen high calibration requires a user supplied source of 100% oxygen.
Note:
If the date is flashing, it signifies the
calibration is past due.
Note: Cap off sample tee if patient gas sample
line is removed from patient breathing
circuit.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...