Part No. 20717 Rev-01
2014-07
1-1
1/ General Information
Indications for Use
• The INOmax DS
IR
(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the
inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric
oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed
injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and
proportional dose of NO. It may be used with most ventilators.
• The INOmax DS
IR
provides continuous integrated monitoring of inspired O
2
, NO
2
, and NO and a
comprehensive alarm system.
• The INOmax DS
IR
incorporates a battery that provides up to six hours of uninterrupted INOMAX delivery in
the absence of an external power source.
• The INOmax DS
IR
includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO
which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patient's breathing
circuit. It may also use the INOblender for backup.
• The target population is controlled by the drug labeling for INOMAX which is currently neonates; refer to the
drug label for specific information. The primary targeted clinical setting is the Neonatal Intensive Care Unit
(NICU) and secondary targeted clinical setting is the transport of neonates.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...