Part No. 20717 Rev-01
2014-07
4-9
10 L/min
1. Use the Pre-Use set-up to connect to
sample gas from the INOblender.
Connect O
2
tubing to the front of the
INOblender.
2.Confirm the injector module is removed
from the Pre-Use set-up and connect the
adapters.
3. On the INOblender, set the INOMAX dose
to 40 ppm and O
2
flow to 10 L/min.
4. Verify NO value on the INOmax DS
IR
.
5. Turn the dose and oxygen flow to zero.
6. Remove the Pre-Use set-up from the
INOblender.
Acceptable
NO Value
32-48 ppm
4
1
2
3
5
5
2
3
40 ppm
4
Note:
• Confirm INOblender inlet hose is
connected to the INOMAX regulator
hose and the Quick-Connect cover
is in place.
• Confirm 50 psig oxygen supply
hose is connected to O
2
inlet fitting
on back of INOblender.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...