"Measure and Display" function
The "
Measure and Display
" function enables the user to check that all XCi channel types configured are
"working". This function will update the channel values based upon the
longest time period that a particular
channel configuration requires to ouput a value. For example, if a depth sensor requires a 25 second "warm-
up", then the values will be updated every 25 seconds. If a Doppler "
Velocity
" channel requires 30 seconds
to calculate a velocity, the the values will be updated every 30 seconds and so on.
1. From the "
Edit configuration settings
" dialogue, click “
Measure and Display
” and the “
View
measurement
” dialogue box will open.
2. Click "
Start
" and the progress bar at the top will begin to advance every second.
3. After all "
Channels
" have been received and/or calculated the "
Channel
" values will be updated.
4. Click “
Stop
” then ”
Close
” to return to the "
Edit configuration settings
" dialogue.
NOTE: "
Total
" channels are not updated in "
Measure and Display
"
For a short video of the procedure click "
Play
"-
MACE XCi User Manual - 212 -
Summary of Contents for HVFlo XCi
Page 1: ......
Page 10: ...MACE XCi User Manual 10...
Page 18: ...MACE XCi User Manual 18...
Page 21: ...MACE XCi User Manual 21...
Page 50: ...MACE XCi User Manual 50...
Page 98: ...MACE XCi User Manual 98...
Page 173: ...MACE XCi User Manual 173...
Page 178: ...MACE XCi User Manual 178...
Page 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Page 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Page 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Page 250: ...MACE XCi User Manual 250...