The User Interface
112
This is useful in a situation where many detectors are in use to stagger the tests to enable
operators to complete them in an orderly fashion.
■
Synch time
This option only appears when the Synchronisation option is selected. It is used to set the exact
time in hours and minutes, using the 24 hour format, to start a PV test.
Tap in the Hour field to display a pop-up numeric keypad screen and enter the number of hours
required. Then tap in the Minutes field and repeat the process.
■
Window
This field is only displayed when either the Interval or Batch Test timing option is selected. It is
used to set a length of time in minutes to successfully complete a PV test, commencing from
the generation of the prompt. When using the Interval Test timing option, the length of time set
cannot be shorter then the time set in the Interval field.
Tap in the field to display a pop-up numeric keypad screen and enter the time in minutes
required.
■
Fault on timeout
This option is displayed for all test type and timing options. When enabled it forces a system
error to be generated if a PV test is not carried out when required.
Tap in the option check box to enable the option.
■
Number of passes
This field is displayed for all test type and timing options. It is used to set the number of passes
required for each test wand type to complete a PV test.
o
Guidance on determining the number of passes to be used
For products that exhibit no product effect and where no Product Registration Photo-eye is
fitted, place each test wand type onto the product sample at the leading edge for the first
pass and then on the trailing edge for the second pass to ensure that the change of reject
timing point still results in the correct rejection of the product. A minimum of two passes
per test wand type is therefore suggested.
For products that exhibit no product effect but where a Product Registration Photo-eye is
fitted, you can place each test wand type anywhere within the product length. In this case
a minimum of one pass per test wand type is sufficient.
For products that exhibit a product effect, regardless of whether a Product Registration
Photo-eye is fitted, place each test wand type onto the product sample at the leading edge
for the first pass, in the middle for the second pass and on the trailing edge for the third
pass to ensure not only that the reject timing is correct for non photo-eye systems but also
to obtain better detection coverage of the product sample itself. A minimum of three
passes per test wand type is therefore suggested.
Tap in the field to display a pop-up numeric keypad screen and enter the number of passes
required.
■
Sample in stats
Tap in the option check box to enable the inclusion of packs passed during the PV test process
in the
Batch Log
statistics.
Summary of Contents for IQ3+
Page 2: ......
Page 36: ...Safety First 26 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 43: ...Installation 33 ...
Page 60: ...Installation 50 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 70: ...About the Metal Detector Range 60 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 218: ...Operating Instructions 208 THIS IS PAGE INTENTIONALLY LEFT BLANK ...
Page 244: ...Maintenance 234 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 248: ...Glossary 238 THIS PAGE IS INTENTIONALLY LEFT BLANK ...