6 oPeration
information
Please refer to the separate
concentration manual for
further details of concentration
measurement.
The converter can also provide the flow velocity based
on a pipe diameter which the customer can freely
program. This value can either be the inner diameter
of the measuring tube (factory default) or the inner
diameter of the process pipe.
6.5.8 Concentration measurement
(Menu C2)
This menu is used to enter the password to activate
concentration measurement (in case the concentration
option is purchased) after the meter is delivered.
6.5.7 Pipe diameter (Menu C1.1.3)
6.5 description of Functions
6.5.9 Flow direction (Menu C1.3.1)
This function allows the operator to set the direction
of flow in relation to the arrow on the housing. If
“Forwards” is selected, the flow direction corresponds
to the “+” arrow and with “Backwards” the flow
direction corresponds to the “-” arrow on the housing.