87
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
Error Type Event Group
Single Event
Description
Actions to Eliminate the
Event
I
I Config: No
Meas.Value
IO A No Meas. Values
The measurement value for the
IO terminals A is not active.
Select another measurement for
IO A in menu C3.2 or activate
the measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
IO B No Meas. Values
The measurement value for the
IO terminals B is not active.
Select another measurement for
IO B in menu C3.3 or activate
the measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
IO C No Meas. Values
The measurement value for the
IO terminals C is not active.
Select another measurement for
IO C in menu C3.4 or activate
the measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
IO D No Meas. Values
The measurement value for the
IO terminals D is not active.
Select another measurement for
IO D in menu C3.5 or activate
the measurement in C5.1 for
diagnosis values or in menu
C2 for concentration values.
Tot 1 No Meas. Values
The measurement value for the
Totalizer 1 is not active.
Select another measurement
for Totalizer 1 in menu C4.1 or
activate the measurement in C1.5
for diagnosis values or in menu
C2 for concentration values.
Tot 2 No Meas. Values
The measurement value for the
Totalizer 2 is not active.
Select another measurement
for Totalizer 2 in menu C4.2 or
activate the measurement in C1.5
for diagnosis values or in menu
C2 for concentration values.
Tot 3 No Meas. Values
The measurement value for the
Totalizer 3 is not active.
Select another measurement
for Totalizer 3 in menu C4.3 or
activate the measurement in C1.5
for diagnosis values or in menu
C2 for concentration values.
Disp. 1 No Meas. Values Measurement value at 1st
measuring page is not active.
Select another measurement for
1st measuring page or activate
the measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.
Disp. 2 No Meas. Values Measurement value at 2nd
measuring page is not active.
Select another measurement for
2nd measuring page or activate
the measurement in C1.5 for
diagnosis values or in menu
C2 for concentration values.
6.6 Status Messages and diagnostic information
legend
Fixed status signal
Variable status signal