69
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
6.5.4 density Calibration (Menu C1.2.1)
Mass flowmeters are calibrated for density in the
factory. Density calibration is based on two calibration
points. In the factory, air and water are used under
reference conditions. The result of this calibration is
Available Options
Option
Explanation
1 Point Calibration
One of the 2 saved calibration points is replaced by customer calibration.
The signal converter decides which of the 2 calibration points is modified.
2nd Calibration Point
Calibration of the second point.
Factory Calibration
The signal converter reactivates the factory settings for the density calibration.
Example of 1 point Calibration with Town Water
Key
Display
Description and Setting
>
A
Quick Setup
Press and hold for 2.5 s, then release the key.
2 x ↑
C1
Setup
↑
C1.2
Density
2 x >
C1.2.1
Density Calibration?
1 Point Calibration
↵
C1.2.1
Density Calib. Product
XXXXXXX
Press ↑
Until
C1.2.1
Density Calib. Product
Town Water
↵
C1.2.1
Density Calibration?
OK
↵
Please Wait
Countdown from 20 s
Density Calib.
Passed
5 x
↵
Save Configuration?
Yes
↵
Display page
saved in the converter electronics and stored in the
factory settings. Regardless, various applications
require maximum accuracy which can only be achieved
with on-site calibration.
6.5 description of Functions