4. eleCtriCal ConneCtionS
4.9 electrical Connection of the inputs and outputs
4.9.1 Field housing, electrical Connection of the inputs and outputs
information
Assembly materials and tools are
not part of the delivery. Use the
assembly materials and tools In
compliance with the applicable
occupational health and safety
directives.
danGer
All work on the electrical
connections may only be carried
out with the power disconnected.
Take note of the voltage data on
the nameplate!
information
For frequencies above 100 Hz,
shielded cables are to be used
in order to reduce effects from
electrical interferences (EMC).
Terminal Compartment for Inputs and Outputs in Field
Housing
1. Open the housing cover.
2. Push the prepared cable through the cable entry and
connect the necessary conductors.
3. Connect the shield if necessary.
4. Close the touch guard.
5. Close the housing cover.
information
Each time a housing cover is
opened, the thread should be
cleaned and greased. Use only
resin-free and acid-free grease.
Ensure that the housing gasket
is properly fitted, clean and
undamaged.
4.9.2 laying electrical Cables Correctly
Protect Housing from Dust and Water
1. Lay the cable in a loop just before the housing.
2. Tighten the screw connection of the cable entry securely.
3. Never mount the housing with the cable entries facing
upwards.
4. Seal cable entries that are not needed with a plug.