50
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
6.4.3 Menu C, Setup
No.
Function
Settings / Descriptions
B3 Information
B3
Information
B3.1
Status Log
Log with date and time of status messages that occurred
B3.2
Status Details
Details about the status messages displayed
B3.3
C Number
CG number, cannot be changed (input/output version)
B3.4
Sensor Electronics
Displays the version of the measuring sensor electronics
B3.5
“Bus interface”
Only appears with PROFIBUS, Modbus, and FF
B3.6
Electronic Revision
For presentation of LC display see Fct. B3.3 and B3.4
B3.7
Sensor Revision
Revision of measuring sensor
B3.8
Change Log
The check sum across all settings is displayed directly below the menu item.
Shows the changes including date, time, and check sum for all settings and parameters in
the converter, regardless of the interface (display, PROFIBUS, etc.) to be used.
No.
Function
Settings / Descriptions
C1 Process Input
C1.1 Zero and Offset
C1.1
Zero and Offset
-
C1.1.1
Zero Calibration
Start zero calibration
Query: calibrate zero?
Setting:
Automatic / Factory Settings /
Manual (display last value; set new value, range: -10 - +10%) /
Break (return with ^ key)
C1.1.2
Zero Add. Offset
Direct setting of zero offset
C1.1.3
Pipe Diameter
Set the pipe diameter in mm to calculate flow velocity
C1.1.4
Flow Correction
Defines additional correction for mass flow;
Range: -100 to +100%
C1.1.5
Zero Calibration Log
Log of last zero calibration including zero point, temperature, date, and time
C1.2 Density
C1.2.1
Density Calib.
Start density calibration
For detailed information refer to Density calibration (menu C1.2.1) on page
67.
C1.2.2
Density Mode
Selection of density mode:
Process (return with ^ key /
Fixed (a fixed value is used for the density (e.g. standard density)) /
Referred (calculates process density based on a reference temperature) /
Standard (calculates standard density based on a reference temperature and the correction
factors)
C1.2.3
Fixed Density Value
Set the fixed value (e.g. standard density) for the density.
Appears only if density mode “Fixed” has been selected in Fct. C1.2.2.
C1.2.3
Density Ref. Temp.
Set the reference temperature for the reference density option
Appears only if density mode “Referred” has been selected in Fct. C1.2.2.
6.4.2 Menu b, test, continued
6.4 Function tables