88
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
Error Type Event Group
Single Event
Description
Actions to Eliminate
the Event
I
I Electr.
Operation Info.
Zero Calibr. Running
A zero calibration is performed.
PROFIBUS: no data
No data exchange via
PROFIBUS.
Set up cyclical
PROFIBUS
communication.
PROFIBUS Baudrate
PROFIBUS searches for the current
Baudrate.
Check PROFIBUS
terminal and connection.
PROFIBUS Passivated
PROFIBUS module is passivated.
Disable PROFIBUS
passivation.
PROFIBUS Clear Data
PROFIBUS Clear Data is active
Start of cyclic PROFIBUS
communication
PROFIBUS Sync Mode
PROFIBUS Sync Mode is active.
Disable Sync Mode.
PROFIBUS Freeze
Mode
PROFIBUS Freeze Mode is active.
Disable Freeze Mode.
Tot 1 Stopped
Totalizer 1 was stopped.
If Totalizer is to continue
counting, select “Yes” in
Fct. C.y.9 (Start
Totalizer).
Tot 2 Stopped
Totalizer 2 was stopped.
Tot 3 Stopped
Totalizer 3 was stopped.
Tot 1 FB4 Threshold
Tot 1 FB4 Threshold exceeded.
Reset Totalizer 1 FB4.
Tot 2 FB5 Threshold
Tot 2 FB5 Threshold exceeded.
Reset Totalizer 2 FB5.
Tot 3 FB6 Threshold
Tot 3 FB6 Threshold exceeded.
Reset Totalizer 3 FB6.
Control In A Active
Control In B Active
Status Out A Active
Status Out B Active
Status Out C Active
Status Out D Active
Fieldbus No Connection Check Fieldbus terminal and connection.
Disp. 1 Overrange
The value in the 1st measurement line of
the display page is limited.
Check setting for 1st
measurement line.
Disp. 2 Overrange
The value in the 2nd measurement line of
the display page is limited.
Check setting for 2nd
measurement line.
Optical Interf. Active
The optical interface is being used. The
optical keys are deactivated.
The keys are ready for
operation again approx.
60 seconds after the
end of the data transfer/
removal of the optical
interface.
6.6 Status Messages and diagnostic information
legend
Fixed status signal
Variable status signal