57
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
No.
Function
Settings / Descriptions
C3._ Pulse Output X
C3._.7
Polarity
Set polarity; the flow direction in C1.3.2 must be noted!
Select:
Both Polarities (plus and minus values are displayed) /
Positive Polarity (display for negative values = 0) /
Negative Polarity (display for positive values = 0) /
Absolute Value (always displays positive, with both negative and positive values)
C3._.8
Low Flow Cutoff
Sets the measurement to “0” for low values
x.xxx±x.xxx%; range: 0.0 - 20%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value
≤
1st value
C3._.9
Time Constant
Range: 000.1 - 100 s
C3._.10
Invert Signal
Select:
Off (activated output: switch closed) /
On (activated output: switch open)
C3._.11
Phase Shift w.r.t. B
Only available when configuring the terminal A or D and only if output B is a pulse or
frequency output. If setting in Fct. C2.5.6 is “Both Polarities”, the phase shift is prefixed by
a symbol, e.g. -90° and +90°
Select:
Off (no phase shift) /0° Phase Shift (between outputs A or D and B, inversion possible) /
90° Phase Shift (between outputs A or D and B, inversion possible) /
180° Phase Shift (between outputs A or D and B, inversion possible)
C3.3.11
Special Functions
This function is only available at the pulse output of terminal B. At the same time, 2 pulse
outputs must be available:
1st output at terminal A or D / 2nd output at terminal B
The B output is operated as a slave output, controlled and set using master output A or D
Select:
Off (no phase shift) /
Phase Shift w.r.t. D or A (slave output is B and master output is D or A)
C3._.12
Information
Serial no. of the I/O board, software version no. and production date of the circuit board
C3._.13
Simulation
Sequence see B1._ Pulse Output X
6.4 Function tables
6.4.3 Menu C, Setup, continued