P
ERSISTENCE
H
ISTOGRAM
S
ETUP
1. In the menu bar, touch
Math
, then
Math Setup
.
2. Touch one of function tabs
F1
through
Fx
The number of math traces available depends on the software
options loaded on your oscilloscope. See specifications..
3. Touch
inside
the
Source1
field and select a source from the pop-up menu.
4. Touch
inside
the
Operator1
field and select
Phistogram
from the
Select Math Operator
menu.
5. Touch
the
Phistogram
tab, then touch inside the Slice Direction field and select
Horizontal
or
Vertical
slice
from the pop-up menu.
6. Touch
inside
the
Slice Center
field and enter a value, using the pop-up keypad.
7. Touch
inside
the
Slice Width
field and enter a value, using the pop-up keypad.
Note:
You can use the front panel Adjust knobs to move the Slice Center line and the Slice Width boundary
lines.
Persistence Trace Range
This math operation has a field where you can enter the percent of the persistence trace population to use in
creating a new waveform.
Persistence Sigma
This math operation has a field where you can enter a scale, measured in standard deviations, by which to create
a new waveform.