4. If applicable, touch the
Pattern
tab. For each input you want to include in the logic pattern, touch inside the
State
field and select a logic state:
Low
or
High
. Select
Don't Care
for all other inputs.
5. Touch
inside
the
Level
data entry field for each input included in the pattern, and enter a voltage level
threshold, using the pop-up keypad.
TV Trigger
TV
triggers provide stable triggering on standard or custom composite video signals. Use them on PAL, SECAM,
or NTSC systems.
TV
T
RIGGER
S
ETUP
1. Touch
the
Trigger
descriptor label.
2. In the Trigger dialog, touch the
TV
trigger button
3. Touch one of these buttons (NTSC or PAL/SECAM) to select a television
standard.
4. Touch
inside
TV Trigger Source
and select the Ext (location of Trigger Out
cable) from the pop-up menu.
Note:
The channel that you have the VT75 Module connected to is
automatically designated TV Signal in the menu.
5. Touch
inside
Polarity
and select positive or negative polarity. This selection is
independent of the polarity selected for Edge trigger (positive) in the "Trigger"
setup dialog.
6. Check
the
Trigger on Any Line
checkbox if you do not want to specify one. In
this case the oscilloscope will trigger on any sync pulse.
7. Touch
inside
# of Fields
and make a selection (1 to 8) from the pop-up menu.
8. Touch
inside
Line
and enter a line number, using the pop-up keypad. The
maximum line number that you can enter depends on the television standard
(NTSC/PAL/SECAM) and the interlace factor selected. For NTSC, line 526 is
the same as line 1; for PAL, line 626 is the same as line 1.
9. Touch
inside
Field
and make a selection from the pop-up menu. The maximum
field number that you can select is limited by your selection in
# of Fields
. If # of
Fields above is set to
1
, this selection is disabled.