![LeCroy WaveRunner 104MXI Operator'S Manual Download Page 21](http://html1.mh-extra.com/html/lecroy/waverunner-104mxi/waverunner-104mxi_operators-manual_1866369021.webp)
O
PERATOR
’
S
M
ANUAL
WRXi-OM-E Rev B
21
•
Power Consumption:
340 watts (340 VA) max., WaveRunner 62Xi: 290 W (290 VA), depending on
accessories installed (probes, PC port plug-ins, etc.); Standby State: 12 watts
•
Physical Dimensions (HWD):
260 mm x 340 mm x 152 mm (10.2 in. x 13.4 in. x 6.0 in.); height
measurement excludes foot pads
•
Weight:
6.95 kg (15.3 lbs.)
Warranty and Service
3-year warranty; calibration recommended yearly
Optional service programs include extended warranty, upgrades, and calibration services.
Environmental Characteristics
T
EMPERATURE
•
Operating:
5 to 40 °C
•
Storage (non-operating):
-20 to +60 °C
H
UMIDITY
•
Operating:
Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50% relative
humidity at 40 °C.
•
Storage (non-operating):
5 to 95% RH (non-condensing) as tested per MIL-PRF-28800F
A
LTITUDE
•
Operating:
Up to 2,000 m
•
Storage (non-operating):
12,192 m (40,000 ft)
R
ANDOM
V
IBRATION
•
Operating:
5 Hz to 500 Hz, overall level: 0.31 g
rms
, 15 minutes in each of 3 orthogonal axes
•
Non-operating:
5 Hz to 500 Hz, overall level: 2.4 g
rms
, 15 minutes in each of 3 orthogonal axes
S
HOCK
•
Functional Shock:
20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of 3
orthogonal axes, 18 shocks total
Certifications
CE Compliant, UL and cUL Listed
CE Declaration of Conformity
The oscilloscope meets requirements of EMC Directive 89/336/EEC for Electromagnetic Compatibility and Low
Voltage Directive 73/23/EEC for Product Safety.
EMC Directive:
•
EN
61326/A3:2003
•
EMC requirements for electrical equipment for measurement, control,
and laboratory use.
Electromagnetic Emissions:
•
EN 55011/A2:2002, Radiated and conducted emissions (Class A)*
•
EN 61000-3-2/A2:2005 Harmonic Current Emissions (Class A)
•
EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers (Pst = 1)
* To conform to Radiated Emissions standard, use properly shielded cables on all I/O terminals.