7. Touch
inside
the
Condition
field in the
ParamCompare
mini-dialog and select a math operator from the pop-
up menu:
.
8. Touch
inside
the
Limit
field and enter a value, using the pop-up numeric keypad. This value takes the
dimension of the parameter that you are testing. For example, if you are testing a time parameter, the unit is
seconds.
If you chose either
WithinDeltaPct
or
WithinDeltaAbs
from the
Condition
menu, touch inside the
Delta
field and enter a value.
Mask Testing
1. Touch
Analysis
in the menu bar, then
Pass/Fail Setup...
in the drop-down menu.
2. Touch
a
Qx
tab; a setup dialog for that position will open.
3. Touch
inside
the
Source1
field and select a source from the pop-up menu.
4. Touch
inside
the
Condition
field in the main dialog and select
Mask Test
.
5. From
the
Test
mini-, make a selection in the
Test is True when
group of buttons:
This selection means, for example, that if you select
All In
the test will be False if even a single waveform
falls outside the mask.
6. From
Show Markers
, choose whether or not to have mask violations displayed.
7. If you are loading a pre-existing mask, touch the
Load Mask
tab, then the
File
button. You can then enter the
file name or browse to its location.
8. If you want to make a mask from your waveform, touch the
Make Mask
tab.
9. Touch
inside
the
Ver Delta
and
Hor Delta
fields and enter boundary values, using the pop-up numeric
keypad.
10. Touch the
Browse
button to create a file name and location for the mask if you want to save it.
11. Touch the
Gate
tab, then enter values in the
Start
and
Stop
fields to constrain the mask to a portion of the
waveform. Or, you can simply touch and drag the Gate posts, which initially are placed at the extreme left and
right ends of the grid.