![LeCroy WaveRunner 104MXI Operator'S Manual Download Page 115](http://html1.mh-extra.com/html/lecroy/waverunner-104mxi/waverunner-104mxi_operators-manual_1866369115.webp)
O
PERATOR
’
S
M
ANUAL
WRXi-OM-E Rev B
115
Parameter
Description
Definition
Notes
Frequency
Frequency: Period of cyclic
signal measured as time
between every other pair of
50% crossings. Starting with
first transition after left cursor,
the period is measured for each
transition pair. Values then
averaged and reciprocal used
to give frequency.
1/period Standard
parameter.
Freq@level
Frequency at a specific level
and slope for every cycle in
waveform.
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels
from noise in data.
Available with JTA2 and
XMAP options.
FWHM
Measures the width of the
largest area histogram peak at
half of the population of the
highest peak.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
FWxx
Measures the width of the
largest area histogram peak at
xx% of the population of the
highest peak.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
Half period
Half period of a waveform.
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels
from noise in data.
Available with JTA2 and
XMAP options.
Hist ampl
Difference in value between the
two most populated peaks in a
histogram.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
Hist base
Value of the left-most of the two
most populated histogram
peaks.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
Hist maximum
Value of the highest (right-
most) populated bin in a
histogram.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
Hist mean
Average or mean value of data
in the histogram.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.