![LeCroy WaveRunner 104MXI Operator'S Manual Download Page 79](http://html1.mh-extra.com/html/lecroy/waverunner-104mxi/waverunner-104mxi_operators-manual_1866369079.webp)
O
PERATOR
’
S
M
ANUAL
WRXi-OM-E Rev B
79
6. Use
the
Auto-Scroll
buttons at the right of the Multi-Zoom dialog to control the zoomed section of your
waveforms:
T
URNING
O
FF
M
ULTI
-Z
OOM
1. In the menu bar, touch
Math
, then
Math Setup...
in the drop-down menu.
2. Touch the Multi-Zoom
On
checkbox to turn off Multi-zoom:
XY Display
Use XY displays to measure the phase shift between otherwise identical signals. You can display either voltage
on both axes or frequency on both axes. The traces must have the same X-axis. The shape of the resulting
pattern reveals information about phase difference and frequency ratio.
XY
D
ISPLAY
S
ETUP
1. In the menu bar, touch
Display
; then touch
Display Setup...
in the drop-down menu.
2. Choose an XY display by touching one of the XY display mode buttons:
3. You have the choice of showing the two waveforms on just the XY grid, or you can also show the input
waveforms on a single or dual grid.
4. Touch
inside
the
Input X
and
Input Y
fields and select your input sources from the pop-up menus. The
inputs can be any combination of channels, math functions, and memory locations.
SAVE AND RECALL
Saving and Recalling Oscilloscope Settings