Non-monotonic Mode
The Non-monotonic search mode looks for edges that cross a threshold more than once between high and low
levels. All events that meet the criteria of slope, hysteresis, and level are presented in a table and highlighted in
the source trace. The value displayed in the table is the difference of the max. and min. of the non-monotonicity.
This can be confirmed with cursors:
The hysteresis value is used to eliminate noise. A non-monotonicity is detected only when its amplitude is greater
than the hysteresis. Therefore, when setting a hysteresis level, set a value that is greater than the amplitude of