Interpolation:
Linear, (sinx)/x
Triggering System
•
Modes:
Normal, Auto, Single, and Stop
•
Sources:
Any input channel, External, Ext/10, or line; slope and level are unique to each source (except line)
•
Coupling Mode:
GND, DC 50 ohms, DC 1 Mohms, AC 1 Mohms
•
Pre-trigger Delay:
0 to 100% of memory size (adjustable in 1% increments or 100 ns)
•
Post-trigger Delay:
10,000 divisions in real time mode; limited at slower time/div settings
•
Holdoff by Time or Events:
1 ns to 20 s or from 1 to 99,999,999 events
•
Internal Trigger Range:
±4.1 div from center (typical)
•
Trigger and Interpolator Jitter:
</= 3 ps
rms
(typical)
•
Maximum Trigger Sensitivity with Edge Trigger (Ch1-4 + external):
44Xi
64Xi
62Xi
104MXi/Xi
204Xi
2 div @ < 400 MHz
1 div @ < 200 MHz
2 div @ < 600 MHz
1 div @ < 200 MHz
2 div @ < 600 MHz
1 div @ < 200 MHz
2 div @ < 1 GHz
1 div @ < 200 MHz
2 div @ < 2 GHz
1 div @ < 200 MHz
•
Maximum Trigger Frequency with SMART Trigger (Ch1-4 + external):
44Xi
64Xi
62Xi
104MXi/Xi
204Xi
400 MHz
@ >/= 10 mV
600 MHz
@ >/= 10 mV
600 MHz
@ >/= 10 mV
1 GHz
@ >/= 10 mV
2 GHz
@ >/= 10 mV
•
Trigger Level DC Accuracy:
±4% of full scale ±2 mV (typical)
•
External Trigger Range:
EXT/10 ±4 V; EXT ±400 mV
Basic Triggers
•
Edge/Slope/Line:
Triggers when the signal meets the slope and level condition.
•
Width:
Triggers on positive or negative pulse widths selectable from 500 ps to 20 s or on intermittent faults
(subject to bandwidth limit of oscilloscope).
•
Pattern:
Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input – 2
Ch+EXT on WaveRunner 62Xi). Each source can be high, low, or don’t care. The High and Low level can be
selected independently. Triggers at start or end of the pattern.
•
State or Edge Qualified:
Triggers on any input source only if a defined state or edge occurred on another
input source. Delay between sources is selectable by time or events.
•
TV:
Provides stable triggering on standard or custom composite video signals. Use them on PAL, SECAM, or
NTSC systems. Optional HDTV Trigger for 1080i, 1080p and 720p formats along with non-standard formats
up to 2000
lines.
SMART Triggers
•
Dropout:
Triggers if the input signal drops out for longer than a selectable time-out between 1 ns and 20 s.
•
Glitch:
Triggers on positive or negative glitches with widths selectable from 500 ps to 20 s or on intermittent
faults (subject to bandwidth limit of oscilloscope).
•
Signal or Pattern Interval:
Triggers on intervals selectable from 1 ns to 20 s.
•
Runt:
Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1
ns and 20 s.
•
Slew Rate:
Activates a trigger when the rising or falling edge of a pulse crosses two threshold levels, an