8-2 Specifications
PLZ-U
8.1
Electrical Speci
fi
cations
Unless speci
fi
ed otherwise, the speci
fi
cations are for the following settings and con-
ditions.
•
The warm-up time is 30 minutes (with current
fl
owing).
•
After warm-up, the PLZ-U must be calibrated correctly in an environment of 23
°
C
±
5
°
C.
•
** % of set denotes ** % of the input voltage, input current, or input power set-
ting.
•
** % of f.s denotes ** % of the rated input voltage, rated input current, or rated
input power.
•
**% of rdg represents denotes ** % of the input voltage, input current, or input
power reading.
*1. Minimum voltage at which the current starts
fl
owing to the PLZ-U is approximately 0.3 V.
(The PLZ-U detects no signal at an input voltage less than or equal to approximately 0.3 V and an input current
less than or equal to approximately 1 % of the range rating. Therefore, if the input voltage is gradually increased
from 0 V, no current will
fl
ow until 0.3 V is exceeded. If a current greater than or equal to 1% of the range rating
starts
fl
owing, the current can
fl
ow at voltages less than equal to 0.3 V.)
*2. Vin: Load input terminal voltage
*3. At a current greater than or equal to (Vin/500 k
Ω
)
*4. Measurement frequency bandwidth: 10 Hz to 1 MHz
*5. Measurement frequency bandwidth: 10 Hz to 20 MHz
Model
PLZ150U
PLZ70UA
Rating
Operating voltage (DC)
*1
1.5 V to 150 V
0 V to 150 V
Current/power
Range
H
30 A / 150 W
15 A / 75 W
M
3 A / 150 W
1.5 A / 75 W
L
300 mA / 45 W
150 mA / 22.5 W
Isolation voltage of the load input terminal
500 VDC
Withstand voltage between load input terminal channels
500 VDC
CC mode
Operating range
Range
H
0 A to 30 A
0 A to 15 A
M
0 A to 3 A
0 A to 1.5 A
L
0 A to 300 mA
0 A to 150 mA
Selectable range
0 % to 105 % of f.s
Resolution
Range
H
2 mA
1 mA
M
0.2 mA
0.1 mA
L
0.02 mA
0.01 mA
Accuracy of setting
Range
H, M, and L
±(0.2 % of set + 0.2 % of f.s) + Vin
*2
/500 k
Ω
Input voltage variation
*3
Range
H
2 mA
M
1 mA
L
0.1 mA
Ripple
rms
*4
3 mA
7.5 mA
p-p
*5
30 mA
50 mA
Summary of Contents for PLZ-30F