PLZ-U
Remote Control 6-29
Sequence commands
Programs a series of steps. For details, see section 5.6, “Sequence Function.”
• If a command includes notations such as
{<conductance>|<cur-
rent>|<power>|<voltage>
the program data corresponding to the selected
program mode is selected.
Creating a sequence
PROG:MEMO
Sets the program memo (comment). Or, queries the current setting.
Up to 11 characters can be speci
fi
ed.
Set command:
PROGram[:SELected]:MEMO "<string>"
Query command:
PROGram[:SELected]:MEMO?
PROG:LOOP
Sets the number of program loops. Or, queries the current setting.
Set command:
PROGram[:SELected]:LOOP <value>
Query command:
PROGram[:SELected]:LOOP?
• Program data
Value
1 to 9998 or 9999 (in
fi
nite loop)
Resolution 1
PROG:LINP/PROG:LOUT
Sets the load on/off condition after the sequence ends. Or, queries the current set-
ting.
Set command:
PROGram[:SELected]:LINPut
{OFF|ON|0|1}
PROGram[:SELected]:LOUTput
{OFF|ON|0|1}
Query command:
PROGram[:SELected]:LINPut?
PROGram[:SELected]:LOUTput?
• Program data
Value
OFF(0)
Turn the load off
ON(1)
Turn the load on
PROG:LVAL
Sets the current value after the end of the program. Or, queries the current setting.
Set command:
PROGram[:SELected]:LVALue
{<conductance>|<current>}
Query command:
PROGram[:SELected]:LVALue?
Returns the current value in response to
PROG:LVAL?
.
Example) When a current value of 9.876 A is speci
fi
ed
Returns 9.876.
NOTE
Summary of Contents for PLZ-30F
Page 32: ...2 18 Installation and Preparation PLZ U ...
Page 58: ...4 14 Basic Operation PLZ U ...
Page 144: ...6 52 Remote Control PLZ U ...
Page 145: ...PLZ U Maintenance 7 1 Chapter 7 Maintenance This chapter explains the PLZ U maintenance 7 7 ...
Page 186: ...A 26 Appendix PLZ U ...
Page 190: ...I 4 Index PLZ U ...
Page 191: ......