PLZ-U
Remote Control 6-13
6.5
SCPI Commands
The SCPI commands below are those that are used on the GPIB or RS-232C inter-
face.
6.5.1
IEEE 488.2 Common Commands
*CLS
Clears all event registers including the status byte, event status, and error queue.
For details on the registers, see section 6.6, “Status Registers.”
Set command:
*CLS
*ESE
Sets the event status register that is counted by the event summary bit (ESB) of the
status byte or queries the current setting.
For details on the event status registers, see section 6.6, “Status Registers.”
Set command:
*ESE
Query command:
*ESE?
Example) When *ESE 16 is transmitted, bit 4 of the event status enable register is
set. Each time the execution error bit (bit 4) of the event status register is
set, the event summary bit (ESB) of the status byte is set.
*ESR
Queries the event status register. This register is cleared when it is read.
For details on the event status registers, see section 6.6, “Status Registers.”
Query command:
*ESR?
*IDN
Queries the manufacturer name, model, and ROM version.
Query command:
*IDN?
The response to
*IDN?
is indicated below.
Example) For PLZ-50F with ROM version 1.00
Returns
KIKUSUI,PLZ-50F,0,1.00
.
*OPC
The PLZ-U does not support asynchronous I/O operation.
When the *OPC command is transmitted, the Operation Complete bit (bit 0) of the
standard event status register is set.
When *OPC? is transmitted, ACII character 1 is returned in the output queue.
For details, see section 12.5.3 of IEEE 488.2-1992.
Set command:
*OPC
Query command:
*OPC?
Summary of Contents for PLZ-30F
Page 32: ...2 18 Installation and Preparation PLZ U ...
Page 58: ...4 14 Basic Operation PLZ U ...
Page 144: ...6 52 Remote Control PLZ U ...
Page 145: ...PLZ U Maintenance 7 1 Chapter 7 Maintenance This chapter explains the PLZ U maintenance 7 7 ...
Page 186: ...A 26 Appendix PLZ U ...
Page 190: ...I 4 Index PLZ U ...
Page 191: ......