6-14 Remote Control
PLZ-U
*RCL
Recalls the contents saved to the memory (0 to 3). The range of parameters that are
con
fi
gured are the same as the range of parameters initialized by the *RST com-
mand. For a description of the commands that are affected, see appendix A.5, “SCPI
Command Reference.”
Set command:
*RCL <value>
• Program data
Value
0 to 3:
Memory number
Resolution 1
*RST
Resets the device to factory default settings. CONFIG number 1 and 9 to 11 in the
con
fi
guration settings are not changed.
The settings after initialization are as follows:
Item
PLZ150U
PLZ70UA
PLZ30F/PLZ50F
Current range
H
H
-
Voltage range
H
H
-
Operation mode
CC
CC
-
CC mode SET value
0.000 A
0.000 A
-
CR mode SET value
0.0 S
0.0 S
-
CV mode SET value
157.50 V
157.50 V
-
Switching function
Off
Off
-
CC switching level
0 A
0 A
-
CR switching level
0 S
0 S
-
Switching frequency
1.000k Hz
1.000k Hz
-
Switching duty cycle
50.0 %
50.0 %
-
Slew rate
2.4 A/
μ
s
1.2 A/
μ
s
-
Soft start
1.0 ms
1.0 ms
-
OCP trip point
33.00 A
16.5 A
-
OPP trip point
165.0 W
82.5 W
-
UVP trip point
Off
Off
-
LOAD OFF TIMER
Off
Off
-
Sequence mode
Off
Off
-
Operation mode
CC
CC
Current range
H RANGE
H RANGE
Step setting
0.000 A
0.000 A
Step execution time
0 ms
0 ms
RPT
1
1
SEQ STEP 1-256
0.000 A, 0 ms
0.000 A, 0 ms
-
SEQ REPEAT
1
1
-
Load
Off
Off
OFF
ENTRY
-
-
CH1
INDEP
Off
Off
-
DISP
Voltage and current
Voltage and current
-
Summary of Contents for PLZ-30F
Page 32: ...2 18 Installation and Preparation PLZ U ...
Page 58: ...4 14 Basic Operation PLZ U ...
Page 144: ...6 52 Remote Control PLZ U ...
Page 145: ...PLZ U Maintenance 7 1 Chapter 7 Maintenance This chapter explains the PLZ U maintenance 7 7 ...
Page 186: ...A 26 Appendix PLZ U ...
Page 190: ...I 4 Index PLZ U ...
Page 191: ......