6-20 Remote Control
PLZ-U
Setting the slew rate
For details on the slew rate setting, see section 5.4, “Setting the Slew Rate.”
• The slew rate can be set to a value outside the speci
fi
cations, but it is actually lim-
ited to a value within the speci
fi
cations of the PLZ-U.
CURR:SLEW
Sets the amount of current change per unit time. Or, queries the current setting.
Set command:
[SOURce:]CURRent:SLEW <slew>
Query command:
[SOURce:]CURRent:SLEW?
• Program data
<slew>
Value
Resolution 0.01
Unit
A/US
Setting protection levels
CURR:PROT
Sets the overcurrent protection. Or, queries the current setting.
Set command:
[SOURce:]CURRent
:PROTection[:LEVel][:OVER]
{<value>|MINimum|MAXimum}
Query command:
[SOURce:]CURRent
:PROTection[:LEVel][:OVER]?
• Program data
Unit
A
Operation mode
Range
PLZ150U
PLZ70UA
CC
H
0.10 to 2.40
0.05 to 1.20
M
0.10 to 0.24
0.05 to 0.12
L (
fi
xed)
24 m
12 m
CR
H
0.10 to 0.24
0.05 to 0.12
M (
fi
xed)
24 m
12 m
L (
fi
xed)
2.4 m
1.2 m
PLZ150U
PLZ70UA
Selectable range
0 to 33.00
0 to 16.5
Resolution
10 m
10 m
NOTE