PLZ-U
Remote Control 6-43
QUEStionable:INSTrument:ISUMmary<n> subregister
Bit <n> subregister of the QUEStionable:INSTrument subregister.
For example, channel 2 (INSTrument2) is QUEStionable:INSTru-
ment:ISUMmary2.
The <n> values that can be used for the PLZ-U are 1 to 5.
Table 6-8
QUEStionable:INSTrument:ISUMmary status register
(STATus:QUEStionable:INSTrument:ISUMmary)
STAT:QUES:INST:ISUM
Queries the event of the QUEStionable:INSTrument:ISUMmary subregister.
Query command:
STATus:QUEStionable:INSTrument
:ISUMmary<n>[:EVENt]?
STAT:QUES:INST:ISUM:COND
Queries the condition of the QUEStionable:INSTrument:ISUMmary subregister.
Query command:
STATus:QUEStionable:INSTrument
:ISUMmary<n>:CONDition?
STAT:QUES:INST:ISUM:ENAB
Sets the QUEStionable:INSTrument:ISUMmary subregister enable, or queries the
current setting.
Set command:
STATus:QUEStionable:INSTrument
:ISUMmary<n>:ENABle
<status-enable>?
Query command:
STATus:QUEStionable:INSTrument
:ISUMmary<n>:ENABle?
Bit
Bit Weight
Bit Name
Description
0
1
OV
Over Voltage
1
2
OC
Over Current
2
4
Not Used
3
8
OP
Over Power
4
16
OT
Over Temperature
5
32
Not Used
6
64
Not Used
7
128
Not Used
8
256
Not Used
9
512
UV
Under Voltage
10
1024
EXT
External Problem
11
2048
REV
Reverse Voltage
12
4096
Not Used
13
8192
Not Used
14
16384
Not Used
15
32768
Not Used
Always zero
Summary of Contents for PLZ-30F
Page 32: ...2 18 Installation and Preparation PLZ U ...
Page 58: ...4 14 Basic Operation PLZ U ...
Page 144: ...6 52 Remote Control PLZ U ...
Page 145: ...PLZ U Maintenance 7 1 Chapter 7 Maintenance This chapter explains the PLZ U maintenance 7 7 ...
Page 186: ...A 26 Appendix PLZ U ...
Page 190: ...I 4 Index PLZ U ...
Page 191: ......