PLZ-U
Remote Control 6-23
Other settings
DISP:MET
Switches the measured value display of the load unit. Or, queries the current setting.
Set command:
DISPlay:METer {0|1|2|3|4}
Query command:
DISPlay:METer?
• Program data
Value
0:
Voltage and current
1:
Voltage and power
2:
Power and current
3:
Elapsed time and alarm
(Alarm display is shown only when an alarm occurs)
4:
Input voltage when the load turns off
FUNC:SST
Sets the soft start time in CC mode. Or, queries the current setting.
The default value is 1 ms.
Set command:
[SOURce:]FUNCtion:SSTart
{0.1MS|1MS|3MS|10MS|30MS|100MS
|300MS|MINimum|MAXimum}
Query command:
[SOURce:]FUNCtion:SSTart?
• Program data
Value
0.0001, 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, MIN, or MAX
Unit
S
HCON:POL
Sets whether to power up with the load turned on. Or, queries the current setting.
Set command:
HCONfiguration:POLoad {OFF|ON|0|1}
Query command:
HCONfiguration:POLoad?
• Program data
Value
ON(1):
Automatically turn the load on at power on
OFF(0): Keep the load turned off at power on