PLZ-U
Remote Control 6-21
CURR:PROT:ACT
Sets whether to turn off the load or limit the current when the OCP trips. Or, queries
the current setting.
Set command:
[SOURce:]CURRent:PROTection
:ACTion {LIMit|TRIP}
Query command:
[SOURce:]CURRent:PROTection:ACTion?
• Program data
Value
LIMit:
Limit the current when the OCP trips.
TRIP:
Turn off the load when the OCP trips.
POW:PROT
Sets the overpower protection. Or, queries the current setting.
Set command:
[SOURce:]POWer:PROTetion[:LEVel]
[:OVER] {<value>|MINimum|MAXimum}
Query command:
[SOURce:]POWer:PROTection[:LEVel]
[:OVER]?
• Program data
Unit
W
POW:PROT:ACT
Sets whether to turn off the load or limit the power when the OPP trips. Or, queries
the current setting.
Set command:
[SOURce:]POWer:PROTection:ACTion
{LIMit|TRIP}
Query command:
[SOURce:]POWer:PROTection:ACTion?
• Program data
Value
LIMit:
Limit the current when the OPP trips.
TRIP:
Turn off the load when the OPP trips.
VOLT:PROT:UND
Sets the undervoltage protection. Or, queries the current setting.
Set command:
[SOURce:]VOLTage:PROTection[:LEVel]
UNDer {<voltage>|MINimum|MAXimum}
Query command:
[SOURce:]VOLTage:PROTection[:LEVel]
UNDer?
VOLT:PROT:STAT
Turns ON/OFF the undervoltage protection. Or, queries the current setting.
Set command:
[SOURce:]VOLTage:PROTection:STATe
{OFF|ON|0|1}
Query command:
[SOURce:]VOLTage:PROTection:STATe?
• Program data
Value
ON(1):
Undervoltage protection on
OFF(0): Undervoltage protection off
PLZ150U
PLZ70UA
Selectable range
0 to 165.0
0 to 82.5
Resolution
100 m
100 m
Summary of Contents for PLZ-30F
Page 32: ...2 18 Installation and Preparation PLZ U ...
Page 58: ...4 14 Basic Operation PLZ U ...
Page 144: ...6 52 Remote Control PLZ U ...
Page 145: ...PLZ U Maintenance 7 1 Chapter 7 Maintenance This chapter explains the PLZ U maintenance 7 7 ...
Page 186: ...A 26 Appendix PLZ U ...
Page 190: ...I 4 Index PLZ U ...
Page 191: ......