Setting Up the Interface
PCR-WE Interface Manual
49
GPIB (option)
The information in this section is valid only when the optional GPIB interface board is
installed.
■
GPIB connection
Connect the product to a PC using a standard IEEE488 cable.
■
GPIB settings
1
Press CONFIG (SHIFT+OPR MODE) > COM-I/F (F1) > GPIB (F4).
The present GPIB settings are displayed.
2
To enable GPIB, press ENABLE (F1). To disable it, press DISABLE (F2).
3
Press ADDRESS (F2) to set the GPIB address.
The factory default GPIB address is 5.
4
Press ESC > APPLY (F5).
To cancel, press CANCEL (F6).
5
Turn the PCR-WE
ʼ
s POWER switch off and then back on.
The settings take effect.
■
GPIB function
Function
Subset
Description
Source handshaking
SH1
Full capability
Acceptor handshaking
AH1
Full capability
Talker
T8
Function available
Listener
L4
Function available
Service request
SR0
No capability
Remote local
RL0
No capability
Parallel polling
PP0
No capability
Device clear
DC0
No capability
Device trigger
DT0
No capability
Controller
C0
No capability
Electrical interface
E1
Open-collector driver