command for the same purpose. This command applies a software trigger to all trig-
ger subsystems, if there are other trigger subsystems in the initiated state, their trig-
ger operations will also be executed at the same time.
How the product operates when triggers are used
When an ABOR command is received, INIT:TRAN is cancelled. The VOLT:TRIG val-
ue does not change.
The following table shows the responses when the voltage is set to 20 V (VOLT 20)
and the voltage set by the trigger is 10 V (VOLT:TRIG 10).
Response
VOLT?
VOLT:TRIG?
Immediately after the command is set
20 V
10 V
After the trigger is sent
10 V
10 V
After a *RST is sent
0 V
0 V
Voltage change VOLT 30 sent before the trigger is sent 30 V
30 V (cancel)