Receiver Power Check
Power accuracy is dependent the accuracy of your power sensor.
After calibration, the reference receivers will show power at the test ports.
Power Level Control and Max Power Level Check
1. Set up two traces, making sure "Statistics" is turned on.
a. For 1-Port and 2-Port: a1,1 and a2,2 (or R1,1 and R2,2 receiver)
b. For 3-Port and 4-port: a3,3 and a4,4 (or R3,3 and R4,4 receiver)
2. Measure the power level at +2 dBm (default). Verify the traces are measuring 2 dBm
(mean), ± 1 dB.
3. Set the power level to 0 dBm. Verify the traces are measuring 0 dBm (mean), ± 1 dB.
4. Set the power level to 13 dBm. Verify the traces are measuring a mean of 13 dBm,±1
dBm.
5. See
for a display example for 2-port, or
for 4 -port.
N5252-90002 User's Guide
41
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...