System Check for N5252A and N5253E
The following system checks will confirm the system is connected and operating
correctly:
Receiver Check
,
Dynamic Range Check
, and
Stability Check
.
Note that all system checks have a 2 dB default power setting.
Refer to the
Help
menu for assistance.
1. Leave the ports open. Select
Preset
.
2. Enter power level +20 dBm
1
to get max test port power output.
3. Select
System > Service > Diagnostics > Receiver Display
4-Port System Receiver Check
Measurement
-20 ±10 dB
Display Window
Table 15
2-Port Receiver Check
A,1
B,2
-20 ±10 dB
-5 ±10 dB
R1,1
R2,2
-5 ±10 dB
4. See
1. Leave the ports open. Select
Preset
.
2. Enter power level +20 dBm
1
to get max test port power output.
3. Select
System > Service > Diagnostics > Receiver Display
Measurement
-20 ±10 dB
Display Window
Table 16
4-Port Receiver Check
A,1
B,2
-20 ±10 dB
-20 ±10 dB
C,3
D,4
-20 ±10 dB
4. See
Figure 24
5.
if receiver level is low, remove DC input to VDI extender module and
reconnect to module.
-5 ±10 dB
R1,1
R2,2
-5 ±10 dB
-5 ±10 dB
R3,3
R4,4
-5 ±10 dB
Receiver Check
2-Port System Receiver Check
1. En20 dBm in the VNA menu will result in approxi13 dBm of output at the waveguide test port.
32
N5252-90002 User's Guide
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...