Alternative System Configurations
Alternative system configurations can be configured with PXI or USB VNA. Refer to the
Banded Millimeter Wave Network Analysis Technical Overview
1. System configuration with P937x (see
2. 4-Port system configuration with Streamline USB VNA (see
3. System configuration with N5252APXI test set and PXI or Streamline VNA with
N5262BWxx standard extenders (see
N5252A/N5253Ex E-Band Bundle
Alternative Configurations
•
The N5252A and N5253E2 allow a 1-port system to be configured. Two 1-port systems
can be configured and operated separately. The VNA application will still show two
ports, even when a second port is not configured. Only use 1-port in calibration and
measurements. The N5253E1 and N5253E3 can be used to make 1-port measurements,
but cannot be configured into two 1-port systems.
•
A 4-port system can be configured with N5253E1 or N5253E3 by adding one more
4-port Streamline VNA and another pair of the frequency extenders.
For alternative system configurations with the P9374/75A VNA, the firmware will need to be
changed to
A.14.10.13
. Refer to the
for drivers, firmware, and software.
The S97551B (Option 551) multiple instruments measurement is required for one of the
Streamline VNA in the system to operate.
Frequency
Extenders for WR15 to WR3.4 mm-Wave Frequency Bands
•
N5252AWxx frequency extenders are available for V, E, W, and D (WR6.5) bands.
•
N5262BWxx frequency extenders with the N5252APXI test set can be configured up
to WR3.4.
•
N5262ACxx waveguide calibration kits are available for each band (C15 - C03).
•
VNAs need to be ≥ 14 GHz for WR15 (V Band) and ≥ 20 GHz for higher frequency
bands.
Streamline
and PXI VNA Requirements
•
Software license S97560B is required for P50xxA to allow configuration for a mm-
wave VNA (not required for P937x).
•
Software license S97551B is required for systems with two Streamline VNAs such as
N5253E2 or 4-port configurations with two P502xAs.
•
You cannot configure a 4-port system with multiple P500xA or P937xA 2-port
Streamline VNAs, as only two USB connections are allowed.
•
You can configure a 4-port system with two P502xA 4-port VNAs.
•
You can use P937xAs for up to WR6.5 band systems and P50xxAs for up to WR3.4
band systems.
•
P502xA cannot be configured into two separate 1-port systems.
N5252-90002 User's Guide
43
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...