Figure 16
N5253E2 Configuration
Cable
Type
1
RF/LO
RF/LO
P5004A
VNA
VDI
Extender #1
Port 1
Port 1
to
Port 2
--
--
Figure
Reference
Table 12
Blue
Red
Blue
N5253E2 Configuration
1. Install the RF/LO and IF cables from Ports 1 and 2 of the P5004A and P5008A to VDI Extenders
#1 and #2. Install RF/LO cable from LO Out of P5004A to LO-I of VDI Extender #2. See
Table 12
and
Figure 16
. Use 11904C adapters on the P5008A.
2. If not already installed, install a 50 Ω load to LO In on P5008A and LO-O on VDI Extender #1.
See
Figure 16
.
3. Connect VDI Extender #1 (LO-I) to VDI Extender #2 (LO-O) with an RF/LO cable. See
Table 12
and
Figure 16
.
4. Connect the power cords to VDI Extenders. See
Figure 16
.
to
to
P5008A
VNA
2
VDI
Extender #2
M
--
RF/R
RF/R
Port 2
M
to
--
Cable Connections for P5004Aand P5008A to VNAX Extenders
Red
IF
IF
LO Out
LO-I
Green
to
--
LO-I
to
LO-O
Green
RF/LO
RF/LO
Black
Control
CTRL S
CTRL M
Black
SMA
LO In
to
LO Out
to
1. RF/LO cables are thicker then the IF cables and are marked RF/LO or IF at the end of the cables.
2. Connections to the P5008A require 11904C adapters.
N5252-90002 User's Guide
25
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...