![Keysight Technologies N5252A User Manual Download Page 24](http://html.mh-extra.com/html/keysight-technologies/n5252a/n5252a_user-manual_1962044024.webp)
Figure 15
N5253E1 Configuration
Cable
Type
1
RF/LO
IF
RF/LO
P5024A
VNA
VDI
Extender #1
Port 1
to
Port 2
RF/R
--
Figure
Reference
Table 11
Cable Connections for P5024A to VNAX Extenders
N5253E1 Configuration
1. Install the RF/LO and IF cables from Ports 1, 2, 3, and 4 of the P5024A to VDI Extenders #1 and
#2. See
Table 11
and
Figure 15
.
2. Install RF/LO cable from LO Out of P5024A (remove 50 Ohm load and set aside for now) and
connect to LO/I of VDI Extender #2. See
Table 11
and
Figure 15
.
3. Install 50 Ω load to LO-O on VDI Extender #1. See
Figure 15
.
4. Connect VDI Extender #1 (LO-I) to VDI Extender #2 (LO-O) with an RF/LO cable. See
Table 11
and
Figure 15
.
5. Connect the power cords to VDI Extenders. See
Figure 15
.
to
to
VDI
Extender #2
RF/R
--
--
M
--
to
M
Port 3
Port 4
LO-I
to
LO-O
--
RF/LO
IF
Control
CTRL M to CTRL S
SMA
LO In to LO Out
Blue
Red
Blue
Red
Green
Black
Black
You can build a 4-port system with the purchase of another N5253E1 system
and duplicating the setup below and connecting CTRL S to CTRL M.
RF/LO
Green
LO Out
--
to
LO-I
1. RF/LO cables are thicker then the IF cables and are marked RF/LO or IF at the end of the cables.
24
N5252-90002 User's Guide
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...