Procedure:
1. Load the cal kit file into your computer. The cal kit file is located on the
USB drive that came with the cal kit. Plug the USB drive into the USB
port on your PC.
2. Select
Cal > Cal Sets & Cal Kits > Cal Kit...
3. Select Import and open the WR12_VDI calibration kit file on the USB
(.ckt or .xkt).
4. Select
Preset
5. Confirm system is in mm-wave mode frequency and setup the number
of points in sweep (default is 201).
6. Select
Cal > Other Cals > Smart Cal
7. Select the box for "Calibrate Power" (see Figure 28 below) if you have a
power meter and E-Band sensor or are using the power data from the
extender USB.
8. Due to the RF and LO multiplication, you may need to calibrate the
VNA source to get the best leveling and power control.
9. Refer to the
Help
menu to complete the Smart Cal.
10. Once the Smart Cal is complete, save the calibration if desired.
Figure 29
Smart Cal GUI
For 1-Port systems, Smart Cal GUI will show two ports (see
). Only select Port 1. Selecting both ports will cause the
calibration to not complete.
N5252-90002 User's Guide
39
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...