Port 4 Configuration - Option 400 Only
Figure 12
Port 4 Configuration
M9374A Module #4 with VDI Extender #4
1. Install the two cables between M9374A Module #4 and VDI Extender #4. See
Table 10
and
Figure 12
.
2. If not already installed, install two 50 Ω loads to LO In and LO Out on M9374A Module #4. If
not already installed, install a reference jumper and a trigger jumper on Modules #3 and #4.
See
Figure 12
.
3. Install a 50 Ω load (use the load removed from Module #1) to VDI Extender #4 (LO-O). See
Figure 12
.
4. Connect the power cord to VDI Extender #4. See
Figure 12
.
Cable
Type
1
RF/LO
IF
M9374A
Module #4
VDI
Extender #4
Port 1
to
Port 2
RF/R
M
Figure
Reference
Table 10
Port 4 Configuration Cable Connections
A
B
to
An RF cable up to 5 meters in length can be used for an LO-O to LO-I connection
between extenders. The M9374A LO Out can only support 1.2 meter cable lengths.
A
B
1. RF/LO cables are thicker then the IF cables and are marked RF/LO or IF at the end of the cables.
N5252-90002 User's Guide
21
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...