Dynamic Range Check for N5252A and N5253E
2-Port System
1. Connect the millimeter wave extender waveguide Ports 1 and 2 together with
supplied captive screws and 5/64 ball driver from the cal kit..
2. Select
Preset
.
3. Set the power level to +20 dBm
1
.
4. Ensure that the following traces are each set up in a display window (turn on
"Statistics" for each trace and refer to the mean value):
a. S12 - Trace
b. S21 - Trace
5. Normalize the S12 and S21 traces.
6. Set IF BW to 10 Hz.
7. Separate the VNA extenders and connect a load to the waveguide test port of each
port.
8. Set the scale reference level to -100 dB in each window.
9. See
Figure 25
below for a display example.
10. Verify the mean value is ≥ -100 dB for each trace.
Figure 25
2-Port Dynamic Range Check with Statistics
1. En20 dBm in the VNA menu will result in approxi13 dBm of output at the waveguide test port.
34
N5252-90002 User's Guide
Summary of Contents for N5252A
Page 1: ...User s Guide Keysight Technologies N5252A and N5253E E Band Vector Network Analyzer Systems ...
Page 5: ...Introduction N5252 90002 User s Guide 5 ...
Page 16: ...System Setup 16 N5252 90002 User s Guide ...
Page 31: ...Figure 22 Installation Calibration Steps N5252 90002 User s Guide 31 ...
Page 33: ...Figure 23 2 Port Receiver Check Figure 24 4 Port Receiver Check N5252 90002 User s Guide 33 ...
Page 48: ...Troubleshooting Maintenance and Support 48 N5252 90002 User s Guide ...
Page 55: ...Safety and Regulatory Information N5252 90002 User s Guide 55 ...