Electronic Calibration Modules Reference Guide 3-13
Operating ECal Module
Validating a Calibration
— The System Verification Procedure is also a method for verifying the
performance of your analyzer.
— Verification kits use accurately known standards with well defined
magnitude and phase response. These kits include precision airlines,
mismatch airlines, and precision fixed attenuators.
— Each verification kit standard is measured precisely by Keysight’s accredited
test laboratory. The measurement data and uncertainty of each standard
are provided in each kit on a disk or USB storage drive.
— Verification kits may be re-certified by Keysight. Contact Keysight for
information about ordering verification kits - see
.
For detailed information on using your analyzer’s built-in System Verification
Procedure, refer to the appropriate service guide or online Help. To use the
internet to view the ENA, PNA, or PXIe online Help, use the steps listed in
“Finding More Information on Performing a Calibration” on page 3-10