![Keysight Technologies 85091D Reference Manual Download Page 134](http://html.mh-extra.com/html/keysight-technologies/85091d/85091d_reference-manual_1961832134.webp)
5-28 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Typical Pin Depth Values
5-
Table 5-43 2.4 mm Connector Pin Depth Characteristics
Device
Typical Pin Depth
in Millimeters
(10
–4
inches)
Measurement Uncertainty
a
in Millimeters
(10
–4
inches)
Observed Pin Depth Limits
in Millimeters
(10
–4
inches)
2.4 mm ECal Module (N4693D) –0.0025 to –0.0152
(–1.0 to –6.0)
+0.0030 to –0.0030
(+1.2 to –1.2)
+0.0005 to –0.0182
(+0.2 to –7.2)
2.4 mm Adapter
0 to –0.0381
(0 to –15)
+0.0030 to –0.0030
(+1.2 to –1.2)
+0.0030 to –0.0411
(+1.2 to –16.2)
a. Approxi2 sigma to –2 sigma of gage uncertainty based on studies performed at the factory using precision digital
gages according to recommended procedures.
Table 5-44 1.85 mm Connector Pin Depth Characteristics
Device
Typical Pin Depth
in Millimeters
(10
–4
inches)
Measurement Uncertainty
a
in Millimeters
(10
–4
inches)
Observed Pin Depth Limits
in Millimeters
(10
–4
inches)
1.85 mm ECal Module (N4694D) –0.0025 to –0.0152
(–1.0 to –6.0)
+0.0030 to –0.0030
(+1.2 to –1.2)
+0.0005 to –0.0182
(+0.2 to –7.2)
1.85 mm Adapter
0 to –0.0381
(0 to –15)
+0.0030 to –0.0030
(+1.2 to –1.2)
+0.0030 to –0.0411
(+1.2 to –16.2)
a. Approxi2 sigma to –2 sigma of gage uncertainty based on studies performed at the factory using precision digital
gages according to recommended procedures.
Table 5-45 7 mm Connector Pin Depth Characteristics
Device
Typical Pin Depth
in Millimeters
(10
–4
inches)
Measurement Uncertainty
a
in Millimeters
(10
–4
inches)
Observed Pin Depth Limits
in Millimeters
(10
–4
inches)
7 mm ECal Module (85091D)
–0.0254 to –0.0508
(–10.0 to –20.0)
+0.0038 to –0.0038
(+1.5 to –1.5)
+0.0216 to –0.0546
(+8.5 to –21.5)
7 mm ECal Module (N4696D)
–0.0025 to –0.0152
(–1.0 to –6.0)
+0.0038 to –0.0038
(+1.5 to –1.5)
+0.0013 to –0.0190
(+0.5 to –7.5)
a. Approxi2 sigma to –2 sigma of gage uncertainty based on studies performed at the factory using precision digital
gages according to recommended procedures.