![Keysight Technologies 85091D Reference Manual Download Page 72](http://html.mh-extra.com/html/keysight-technologies/85091d/85091d_reference-manual_1961832072.webp)
3-6 Electronic Calibration Modules Reference Guide
Operating ECal Module
Setting Up a Calibration
3-
Calibration Types Available
The following table contains the calibration types and advanced settings
available for all compatible network analyzers:
Table 3-3
Calibration Types and Advanced Settings Available
Calibration Types
Model
PNA
PXIe
Streamline
VNA
ENA
FieldFox
a
1-Port Reflection
Yes
Yes
Yes
Yes
Yes
Full 2-Port
Yes
Yes
Yes
Yes
Yes
Full N-Port where N>2
Yes
Yes
Yes
Yes
No
Enhanced Response
Yes
Yes
Yes
Yes
Yes
QSOLT
Yes Yes
Yes
No
Yes
Advanced Settings
Adapter Removal
Yes
Yes
Yes
No
No
Flush (zero length) THRU
Yes
Yes
Yes
No
Yes
Unknown THRU
Yes
Yes
Yes
Yes
Yes
User-Characterization
Yes
Yes
Yes
Yes
Yes
b
a. ECal calibrations are not supported on the N9912A.
b. FieldFox user-characterizations must be performed on a PNA or ENA. Learn more about User
Characterization at the PNA Help website:
http://na.support.keysight.com/pna/help/latest/S3_-