![Keysight Technologies 85091D Reference Manual Download Page 68](http://html.mh-extra.com/html/keysight-technologies/85091d/85091d_reference-manual_1961832068.webp)
3-2 Electronic Calibration Modules Reference Guide
Operating ECal Module
Setting Up a Calibration
3-
Table 3-2
8509xD, N443xD and N469xD Status LED Operation
Status LED Color
Description
Off
— ECal is not connected to an
instrument or is malfunctioning.
— See also, Figure
Orange
— ECal is warming up.
— ECal connected and functioning, but
is not communicating with the driver
software.
— See also, Figure
Steady Green
— ECal is connected, functioning,
warmed up, and communicating
normally.
a
Alternating Green/Orange Flashing
— ECal is too cold and unable to
regulate internal temperature to
setpoint.
— Performance may be degraded
beyond specifications.
— See also, Figure
Alternating Green/Red Flashing
— ECal is too hot and unable to
regulate internal temperature to
setpoint.
— Performance may be degraded
beyond specifications.
— See also, Figure
a. Status LED briefly flashes orange when the ECal is first connected to an instrument while the
driver software loads.