![Keysight Technologies 85091D Reference Manual Download Page 115](http://html.mh-extra.com/html/keysight-technologies/85091d/85091d_reference-manual_1961832115.webp)
Electronic Calibration Modules Reference Guide 5-9
Specifications and Characteristics
Electrical Specifications
N4431D (7-16)
The electrical performance in the following table applies to N4431D (if any of
the ports consist of 7-16 connector).
N4431D (4.3-10)
The electrical performance in the following table applies to N4431D (if any of
the ports consist of 4.3-10 connector).
Table 5-14 Electrical Performance for N4431D (m/f 7–16 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz 5 GHz to 7.5 GHz
Directivity (dB)
45
50
47
Source Match (dB) 36
42
39
Reflection Tracking
(±dB)
0.10
0.06
0.09
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance
indicated in this table.
Table 5-15 Electrical Performance for N4431D (m/f 4.3–10 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz 3 GHz to 9 GHz
9 GHz to 12 GHz
Directivity (dB)
45
50
47
35
Source Match (dB) 36
42
39
37
Reflection Tracking
(±dB)
0.10
0.06
0.09
0.10
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance
indicated in this table.