Remote programming output-off states quick reference
The content of the following table is a quick reference of commands for programming output-off states
from a remote interface.
Output-off state programming quick reference
Command Description
smua.source.offmode = smua.OUTPUT_NORMAL
Selects normal output-off state.
smua.source.offmode = smua.OUTPUT_HIGH_Z
Selects high-impedance output-off
state.
smua.source.offmode = smua.OUTPUT_ZERO
Selects zero output-off state.
smua.source.offfunc = smua.OUTPUT_DCVOLTS
Sets 0 V output with current limit
set by
smua.source.offlimiti
attribute.
smua.source.offfunc = smua.OUTPUT_DCAMPS
Sets 0 A output with voltage limit
set by
smua.source.offlimitv
attribute.
smua.source.offlimiti = iValue
Sets current limit in normal
output-off state.
smua.source.offlimitv = vValue
Sets voltage limit in normal
output-off state.
USB storage overview
The Keithley Instruments Model 2651A High Power System SourceMeter
®
Instrument includes a USB
port on the front panel. To store scripts and to transfer files from the instrument to the host PC, insert
a USB flash drive into the USB port.
•
For additional information about saving reading buffers to the USB flash drive, see
•
For additional information about storing and loading scripts to and from the USB flash drive, see
•
For additional information about file I/O, see File I/O.
•
For additional information about saving user setups, see
(on page 2-39).
Connecting the USB flash drive
The Model 2651A supports flash drives that comply with USB 1.0 and 2.0 standards. You can save
data to the USB flash drive from the front panel, or you can create a script to save data to the USB
flash drive.
To connect the USB flash drive, plug the USB flash drive into the USB port located on the
instrument's front panel (see the figure below).