Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 8: Troubleshooting guide
2651A-901-01 Rev. A / March 2011
8-5
Error summary
Error number
Error level
Error Message
2408
RECOVERABLE
TSPnet write failure, aborted
2409
RECOVERABLE
TSPnet write failure, timeout
2410
RECOVERABLE
TSPnet max connections reached
2411
RECOVERABLE
TSPnet connection failed
2412
RECOVERABLE
TSPnet invalid termination
2413
RECOVERABLE
TSPnet invalid reading buffer table
2414
RECOVERABLE
TSPnet invalid reading buffer index range
2415
RECOVERABLE
TSPnet feature only supported on TSP connections
2416
RECOVERABLE
TSPnet musty specify both port and init
2417
RECOVERABLE
TSPnet disconnected by other side
4900
RECOVERABLE
Reading buffer index %s is invalid
4901
RECOVERABLE
The maximum index for this buffer is %d
4903
RECOVERABLE
Reading buffer expired
4904
SERIOUS
ICX parameter count mismatch, %s (Line #%d)
4905
SERIOUS
ICX parameter invalid value, %s (Line #%d)
4906
SERIOUS
ICX invalid function id, %s (Line #%d)
5001
FATAL
SMU is unresponsive
5003
SERIOUS
Saved calibration constants corrupted
5004
SERIOUS
Operation conflicts with CALA sense mode
5005
SERIOUS
Value too big for range
5007
SERIOUS
Operation would exceed safe operating area of the instrument
5008
SERIOUS
Operation not permitted while output is on
5009
SERIOUS
Unknown sourcing function
5010
SERIOUS
No such SMU function
5011
SERIOUS
Operation not permitted while cal is locked
5012
SERIOUS
Cal data not saved - save or restore before lock
5013
SERIOUS
Cannot save cal data - unlock before save
5014
SERIOUS
Cannot restore cal data - unlock before restore
5015
SERIOUS
Save to cal set disallowed
5016
SERIOUS
Cannot change cal date - unlock before operation
5017
SERIOUS
Cannot change cal constants - unlock before operation
5018
SERIOUS
Cal version inconsistency
5019
SERIOUS
Cannot unlock - invalid password
5021
SERIOUS
Cannot restore default calset. Using previous calset
5022
SERIOUS
Cannot restore previous calset. Using factory calset
5023
SERIOUS
Cannot restore factory calset. Using nominal calset
5024
SERIOUS
Cannot restore nominal calset. Using firmware defaults
5025
SERIOUS
Cannot set filter.count > 1 when measure.count > 1
5027
SERIOUS
Unlock cal data with factory password
5028
SERIOUS
Cannot perform requested operation while source autorange is
enabled
5029
SERIOUS
Cannot save without changing cal date and cal due values
5032
RECOVERABLE
Cannot change this setting unless buffer is cleared
5033
RECOVERABLE
Reading buffer not found within device
5038
RECOVERABLE
Index exceeds maximum reading
5040
RECOVERABLE
Cannot use same reading buffer for multiple overlapped
measurements
5041
SERIOUS
Output Enable not asserted
5042
SERIOUS
Invalid while overlapped measure
5043
SERIOUS
Cannot perform requested operation while voltage measure
autorange is enabled