Appendix B: Calibration
Model 2651A High Power System SourceMeter® Instrument Reference Manual
B-22
2651A-901-01 Rev. A / March 2011
Calibration commands
Command Description
smua.measure.calibratev(range,
cp1measured, cp1reference,
cp2measured, cp2reference)
Calibrate voltage measure range*:
±range (measurement range to calibrate).
cp1measured (Model 2651A measured value for cal. point 1).
cp1reference (reference measurement for cal. point 1).
cp2measured (Model 2651A measured value for cal. point 2).
cp2reference (reference measurement for cal. point 2).
smua.source.calibratei(range,
cp1expected, cp1reference,
cp2expected, cp2reference)
Calibrate current source range*:
±range (range to calibrate).
cp1expected (source value programmed for cal. point 1).
cp1reference (reference measurement for cal. point 1).
cp2expected (source value programmed for cal. point 2).
cp2reference (reference measurement for cal. point 2).
smua.source.calibratev(range,
cp1expected, cp1reference,
cp2expected, cp2reference)
Calibrate voltage source range*:
±range (range to calibrate).
cp1expected (source value programmed for cal. point 1).
cp1reference (reference measurement for cal. point 1).
cp2expected (source value programmed for cal. point 2).
cp2reference (reference measurement for cal. point 2)
smua.contact.calibratelo(
cp1measured, cp1reference,
cp2measured, cp2reference)
Calibrate the low/sense low contact check measurement.
cp1measured (value measured by SMU for cal. point 1).
cp1reference (reference measurement for cal. point 1).
cp2measured (value measured by SMU for cal. point 2).
cp2reference (reference measurement for cal. point 2).
smua.contact.calibratehi(
cp1measured, cp1reference,
cp2measured, cp2reference)
Calibrate the high/sense high contact check measurement.
cp1measured (value measured by SMU for cal. point 1).
cp1reference (reference measurement for cal. point 1).
cp2measured (value measured by SMU for cal. point 2).
cp2reference (reference measurement for cal. point 2)
* Calibration point 1 should be performed at approximately 0% of range; calibration point 2 should be performed at
approximately 90% of range. See
(on page B-19) for calibration points.
Calibration adjustment procedure
Use the following procedure to perform remote calibration by sending commands over the IEEE-488
bus, RS-232 port, or LAN. The remote commands and appropriate parameters are separately
summarized for each step.
Step 1. Prepare the Model 2651A for calibration
A. Connect the Model 2651A to the controller IEEE-488 interface, RS-232 port, or LAN using an
appropriate interface cable.
B. Turn on the Model 2651A and the test equipment, and allow them to warm up for at least two
hours before performing calibration.
C. Make sure the IEEE-488, RS-232, or LAN interface parameters are set up properly (to configure
the interface, press the
MENU
key, and then select
RS232
,
LAN
, or
GPIB
, as applicable).