Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-261
Bit Value
Description
B0
Not used
Not applicable.
B1
status.measurement.reading_overflow.SMUA
Set bit indicates that an overflow reading has
been detected.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event reading overflow summary enable register, set
status.measurement.reading_overflow.enable =
status.measurement.reading_overflow.SMUA
.
In addition to the above constant,
measurementRegister
can be set to the numeric equivalent of the bit to set.
Example
status.measurement.reading_overflow.enable =
status.measurement.reading_overflow.SMUA
Sets the SMU A bit of the
measurement reading overflow
summary enable register using a
constant.
Also see
(on page E-23)
status.measurement.voltage_limit.*
This attribute contains the measurement event voltage limit summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
2 (All bits set)
Usage
measurementRegister
= status.measurement.voltage_limit.condition
measurementRegister
= status.measurement.voltage_limit.enable
measurementRegister
= status.measurement.voltage_limit.event
measurementRegister
= status.measurement.voltage_limit.ntr
measurementRegister
= status.measurement.voltage_limit.ptr
status.measurement.voltage_limit.enable =
measurementRegister
status.measurement.voltage_limit.ntr =
measurementRegister
status.measurement.voltage_limit.ptr =
measurementRegister
measurementRegister
The measurement voltage limit summary register status. A zero (0) indicates no
bits set (also send 0 to clear all bits); the only valid value other than 0 is 2