Section 3: Functions and features
Model 2651A High Power System SourceMeter® Instrument Reference Manual
3-34
2651A-901-01 Rev. A / March 2011
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smu
X
.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smu
X
.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[
N
].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object
N
display.trigger.EVENT_ID
Occurs when the TRIG key on the front panel is
pressed.
trigger.EVENT_ID
Occurs when a *TRG command is received on the
remote interface.
(GPIB only) Occurs when a GET bus command is
received.
(VXI-11 only) Occurs with the VXI-11 command
device_trigger
. Reference the VXI-11 standard
for additional details on the device trigger operation.
trigger.blender[
N
].EVENT_ID
Occurs after a collection of events is detected.
trigger.timer[
N
].EVENT_ID
Occurs when a delay expires.
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Using the remote trigger model
The source-measure unit (SMU) in the Model 2651A has a remote trigger model that supports a wide
range of triggering features for source sweeps, triggered measurements, and pulse actions.
Measurements using the trigger model can be made synchronously with sourcing actions, or they can
be made asynchronously. The following figures graphically illustrate both modes of the remote trigger
model.