Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-189
smuX.contact.threshold
This attribute stores the resistance threshold for the
smu
X
.contact.check()
function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Not saved
50
Usage
rValue
= smu
X
.contact.threshold
smu
X
.contact.threshold =
rValue
rValue
The resistance above which contact check should fail (measured in ohms).
X
SMU channel (for example,
smua.contact.threshold
applies to SMU Channel
A).
Details
The threshold should be set to less than 1 k
Ω
.
Example
smua.contact.threshold = 5
Set the contact check threshold for SMU Channel A to 5
Ω
.
Also see
(on page 7-186)
(on page 7-209)
(on page 2-45)
(on page 2-37)
smuX.makebuffer()
This function creates a reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
bufferVar
= smu
X
.makebuffer(
bufferSize
)
bufferVar
The created reading buffer.
X
Source-measure unit (SMU) channel (for example,
smua.makebuffer()
applies
to SMU channel A).
bufferSize
Maximum number of readings that can be stored.